Available Models | Volume | Accuracy | Precision | ||
---|---|---|---|---|---|
%D | µl | %CV | µl | ||
p20 GEN2 | 1 | ±15% | 0.15µL | ±5% | 0.05µL |
10 | ±2% | 0.2µL | ±1% | 0.1µL | |
20 | ±1.5% | 0.3µL | ±0.8% | 0.16µL | |
p300 GEN2 | 20 | ±4% | 0.8µL | ±2.5% | 0.5µL |
150 | ±1% | 1.5µL | ±0.4% | 0.6µL | |
300 | ±0.6% | 1.8µL | ±0.3% | 0.9µL | |
p1000 GEN2 | 100 | ±2% | 2.0µL | ±1% | 1µL |
500 | ±1% | 5.0µL | ±0.2% | 1µL | |
1000 | ±0.7% | 7.0µL | ±0.15% | 1.5µL |
Available Models | Volume | Accuracy | Precision | ||
---|---|---|---|---|---|
%D | µl | %CV | µl | ||
P20 GEN2 | 1 | ±20% | 0.2µL | ±10% | 0.1µL |
10 | ±3% | 0.3µL | ±2% | 0.2µL | |
20 | ± 2.2% | 0.44µL | ±1.5% | 0.3µL | |
P300 GEN2 | 20 | ±10% | 2.0µL | ±4% | 0.8µL |
150 | ±2.5% | 3.75µL | ±0.8% | 1.2µL | |
300 | ±1.5% | 4.5µL | ±0.5% | 1.5µL |